E-Case report

Tumor Fracture T5
Posterior fixation T2–T8
Dr. David Giber

Pre OP

Clinical Case –

Tumor Fracture T5

Posterior fixation T2 – T8

Neo ADVISE™ Trauma module

Dr. David Giber
Orthopedic surgeon
AP-HP – Hôpital Universitaire Henri Mondor
Creteil, Paris – France

Planned Surgery:

1.General Anestesia
2.Fracture reduction T5
3.Posterior fixation  T2 – T8
4.Open approach – Midline incision
5.Neo ADVISE™ application to be used

Pre OP CT sagittal

marco grosso pre operation representation

Intra OP

Navigation of the jamshidi guide wires

Decompression & Laminectomy performed

Posterior fixation with Neo Pedicle Screw System™

Totally 12 pedicle screws placed over the vertebral levels T2 – T8

T2 – 2 x Ø5.0 x 35 mm – polyaxial
T3 – 2 x Ø5.0 x 35 mm – monoplanar
T4 – 2 x Ø5.0 x 35 mm – monoplanar
T6 – 2 x Ø5.0 x 35 mm – monoplanar
T7 – 2 x Ø4.5 x 35 mm – monoplanar
T8 – 2 x Ø5.0 x 40 mm – polyaxial

Cement augmentation of the pedicle screws in levels T2, T3 and T7, T8

Usage of Neo ADVISE™ trauma module

Scanning of the guide towers

Usage of Neo ADVISE™ trauma module

ADVISE™ trauma module
Rod selection: 160 mm straight
Pre OP kyphosis: 19°
Correction of 8° to have a residual Post OP kyphosis of 11°

Patient specific rod contouring integrating the targeted 8° correction and visulized in the iPad screen

Rod insertion without any constraint, this is beneficial considering the compromised bone quality of this patient.

Post OP

Radiographs after final fixation of the posterior construction

Published with the approval of
Dr. David Giber